RsrcName indicates the resource name for a VISA instrument. When you create a VISA object, RsrcName is automatically assigned the value specified in the visa function.
The resource name is a symbolic name for the instrument. The resource name you supply to visa depends on the interface and has the format shown below. The components in brackets are optional and have a default value of 0, except port_number, which has a default value of 1.
Interface | Resource Name |
|---|---|
VXI | VXI[chassis]::VXI_logical_address::INSTR |
GPIB-VXI | GPIB-VXI[chassis]::VXI_logical_address::INSTR |
GPIB | GPIB[board]::primary_address[::secondary_address]::INSTR |
TCPIP | TCPIP[board]::remote_host[::lan_device_name]::INSTR |
RSIB | RSIB::remote_host::INSTR |
Serial | ASRL[port_number]::INSTR |
USB | USB[board]::manid::model_code::serial_No[::interface_No]::INSTR |
If you change the BoardIndex, ChassisIndex, InterfaceIndex, LANName, LogicalAddress, ManufacturerID, ModelCode, Port, PrimaryAddress, RemoteHost, SecondaryAddress, or SerialNumber property value, RsrcName is automatically updated to reflect the change.
Usage | VISA-GPIB, VISA-VXI, VISA-GPIB-VXI, VISA-serial |
Read only | Always |
Data type | String |
The value is defined when the instrument object is created.
To create a VISA-GPIB object associated with a GPIB controller with board index 0 and an instrument with primary address 1, you supply the following resource name to the visa function.
vg = visa('ni','GPIB0::1::INSTR');To create a VISA-VXI object associated with a VXI chassis with index 0 and an instrument with logical address 130, you supply the following resource name to the visa function.
vv = visa('agilent','VXI0::130::INSTR');To create a VISA-GPIB-VXI object associated with a VXI chassis with index 0 and an instrument with logical address 80, you supply the following resource name to the visa function.
vgv = visa('agilent','GPIB-VXI0::80::INSTR');To create a VISA-serial object associated with the COM1 serial port, you supply the following resource name to the visa function.
vs = visa('ni','ASRL1::INSTR');BoardIndex, ChassisIndex, InterfaceIndex, LANName, LogicalAddress, ManufacturerID, ModelCode, Port, PrimaryAddress, RemoteHost, SecondaryAddress, SerialNumber